RS232 Receiver Time-Out

The unit stays awake to receive data until the RS232 Receiver Time-Out expires. A manual or serial trigger resets the time-
out. When an RS232 receiver is sleeping, a character may be sent to wake up the receiver and reset the time-out. A trans-
action on the CTS line will also wake up the receiver. The receiver takes 300 milliseconds to completely come up. Change
the RS232 receiver time-out by scanning the bar code below, then scanning digits from the inside back cover of this man-
ual, then scanning Save. The range is 0 to 300 seconds. Default = 0 seconds (no time-out - always on).

RS232 Handshaking

RS232 Handshaking allows control of data transmission from the scanner using software commands from the host device.
When RTS/CTS is turned Off, no data flow control is used.
Flow Control, No Timeout: The scanner asserts RTS when it has data to send, and will wait indefinitely for CTS to be
asserted by the host.
Two-Direction Flow Control: The scanner asserts RTS when it is OK for the host to transmit. The host asserts CTS
when it is OK for the device to transmit.
Flow Control with Timeout: The scanner asserts RTS when it has data to send and waits for a delay (see
Timeout
on page 2-21) for CTS to be asserted by the host. If the delay time expires and CTS is not asserted, the device
transmit buffer is cleared and scanning may resume.
Default = RTS/CTS Off.
Flow Control, No Timeout
Flow Control with Timeout
2 - 20
RS232 Receiver Time-Out
7 Data, 2 Stop, Parity Mark
RS232
Two-Direction Flow Control
* RTS/CTS Off