9 Data output RS 232 C

9.1 Technical data

8-bit ASCII Code
1 start bit, 8 data bits, 1 stop bit, no parity bit
Baud rate selectable at 1200, 2400, 4800, 9600 and 19200 Baud
Miniature plug-in necessary (9 pole D-Sub)
For operation with interface faultless operation is only ensured with the correct
KERN – interface cable (max. 2m)

9.2 Pin allocation of the balance output socket (front view)

5

9.3 Interface parameter

General Information
The previous condition for the data transfer between balance and a peripherical
device (e.g. printer, PC) is that the appliances are set to the same interface
parameters (e.g. baud rate, transfer mode).

9.3.1 Data transfer mode

With the weighing scale switched off, press ON/OFF and TARE-key
simultaneously until the first function "Unit" appears.
Repeatedly press the 10x-key until "PR" appears.
Confirm by pressing the PRINT-key. Current setting appears.
Now you can use the 10x-key to select among the different settings
(see chpt. 8.3)
Changed setting will be imported by pressing the PRINT key. The balance
returns automatically into weighing mode.
26
1
9
6
Pin 2:
Transmit data
Pin 3:
Receive data
Pin 5:
Signal ground