Interface Rs 232 C; Rs 232 C Data Output Via Interface Rs 232 C; There Are 4 Methods For The Data Output Via Rs 232 C - KERN KB Series Operating Instructions Manual

Electronic precision and platform balances
Table of Contents

Advertisement

7.3 Interface RS 232 C

Technical Data
8-bit ASCII Code
1 start bit, 8 data bits, 1 stop bits, no parity bit
Baud rate adjustable to 2400, 4800 and 9600 baud (default), 19200 baud
Sup-D- 9 pol. is necessary .
When working with an interface correct operation is secured only if the corresponding KERN-interface-
cable (max. 2m) is used.
Description of the jack
Sup-D- 9 pol.
Baudrate
The Baud rate for the data transfer is adjusted with the MODE-key. The following example demonstrates
how to set the Baud rate 4800.
Select Baud rate
1. Press MODE-key repeatedly until "PRINTER" is displayed.
2. Press YES-key.
3. Press MODE-key repeatedly until the desired Baud rate appears (for
instance 4800 Baud).
4. Press YES-key to select 4800 Baud. The tick-mark (X) confirms the new
setting.
5. Press MODE-key repeatedly until the balance displays in grams again,
or press tare key.

7.4 RS 232 C Data output via interface RS 232 C

RS 232 C Data output via interface RS 232 C
General information
As a condition for the data transfer between the balance and a peripheral device (for instance printer, PC ...)
both devise have to be set on the same interface parameter (for instance baud rate, parity ...).

7.4.1 There are 4 methods for the data output via RS 232 C

Data output via PRINT-Key
The printing process can be released by the PRINT-key. In this case the settings AUTOPRINT and AUT-
PRINT PC should be deselected.
AUTOPRINT (Data output, after having loaded the balance)
The setting AUTOPRINT is in the PRINTER-routine, and there it can be selected or deselected. When
AUTOPRINT is active the actual weighing value will be sent via the RS 232 interface when the balance has
been unloaded and then loaded after having achieved the stability.
AUTOPRINT PC (Continuous data output)
The setting AUTPRINT PC is in the PRINTER-routine, and there it can be selected or deselected. When
AUTOPRINT PC is active the actual weighing values will be sent continuously via the RS 232 interface.
Data output by transfer of remote controls
The following functions can be released by the remote controls that will be transferred as ASCII signs to the
balance.
t
Tare.
w
a weighing value (or unstable) is sent via RS 232 interface.
s
a stable weighing value is sent via RS 232 interface.
If the balance receives the command w or s, it acts without printing delay.
40
5
1
9
6
Pin 2:
Transmit data
Pin 3:
Receive data
Pin 5:
Signal ground
572/573/KB/DS/FKB-BA-defsi-0651
Display
PRINTER?
2400 Baud
4800 Baud
4800 Baud X
0,0 g