8.4.4 Selection of the adjustment weight

For the model ranges KERN CDE and CME the adjustment weight can be selected
from three or four preset nominal values (c.1/3; 2/3; max or c.1/4; 1/2; 3/4; max) (see
also table 1 below, default settings are highlighted in grey). In order to achieve high-
quality weighing results in the sense of the measuring technology, it is recommended
to select the nominal value as high as possible.
With the weighing scale switched off, press ON/OFF and TARE-key
simultaneously until the first function "Unit" appears.
Repeatedly press the 10x-key until "CAL" appears.
Confirm by pressing the PRINT-key. Current setting appears.
Use the 10x-key, to select among the preset nominal values (see
tab. 1)
Confirm your selected setting by pressing the PRINT-key.
CME 100-3
CME 300-2
20 g
100 g
50 g
200 g
100 g
300 g
CDE 35K1
CDE 35K1L
5 kg
5 kg
10 kg
10 kg
20 kg
20 kg
30 kg
30 kg

8.4.5 Reset to factory setting

With this function the manual changes of the menu settings are reset to factory
setting.
With the weighing scale switched off, press ON/OFF and TARE-key
simultaneously until the first function "Unit" appears.
Repeatedly press the 10x-key until "rSt" appears.
Confirm by pressing the PRINT-key. Current setting appears.
Now you can use the 10x-key to select among the settings below:
„rSt
no": No reset to default setting
„rSt
yes": Reset to default setting
Confirm your selected setting by pressing the PRINT-key.
CDE/CME/CDEE-BA-e-1014
CME 1000-2 CME 3000-1 CME 6000-1
200 kg
1 kg
500 kg
2 kg
1000 kg
3 kg
CDE 60K2
CDE 60K2L
20 kg
20 kg
40 kg
40 kg
60 kg
60 kg
2 kg
5 kg
6 kg
CDE 150K5 CDE 150K5L
50 kg
50 kg
100 kg
100 kg
150 kg
150 kg