For the internal channel, use equation 45
For the 100 ohm probes, use equation 86
For the 1000 ohm probes, use equation 87
For results in resistance (ohms):
For the internal channel, use equation 45
For the 100 ohm probes, use equation 84
For the 1000 ohm probes, use equation 85
NOTE: Even equation numbers are for 100 W probes and odd equation
numbers are for 1000 W probes.
4.
Adjust the remaining setup options as desired then save the setup.
1.13.3
Specification of RTD Channels
The following specifications are for:
4-Wire Platinum 100 Ohm @ 0° C & TCR = 0.00385
4-Wire Platinum 1000 Ohm @ 0° C & TCR = 0.00385
Description:
+SIG Input Impedance:
-SIG Input Impedance:
Excitation Current
Temp. Range
(100 & 1000 W probes):
Accuracy ** (100 & 1000 W):
Resolution (100 W & 1000 W): ± 0.14°C (±0.25°F)
Resistance Range (100 W):
Accuracy ** (100 W):
Resistance Range (1000 W):
Accuracy ** (1000 W):
Resolution (100 W):
Resolution (1000 W):
* For internal channel specifications, please see
SmartReader Plus loggers
** Temperature compensated for the range of -20 to 80°C (-4 to 176°F)
Copyright © 2010 ACR Systems Inc. All Rights Reserved.
SmartReader Plus Data Logger
Value:
>1 Meg. ohms
>1 Meg. ohms
<=1mA
-230 to 300°C (-382 to 572°F)
± 0.3°C (0.54°F) + resolution or
± 0.06 % full scale + resolution
5 to 210 W
± 0.13 W + resolution or
± 0.06 % full scale + resolution
50 to 2100 W
± 1.3 W + resolution or
± 0.06 % full scale + resolution
± 0.05 W
± 0.5 W
Common Specifications
66
for