1.12
SmartReader Plus 9 Pulse Logger
The SmartReader Plus 9 is a two-channel multi-purpose counter and logger. With it you can
monitor a wide variety of measurement parameters.
1.12.1
Description
Featuring two input channels configured for easy interface with common switch and
transducer outputs, the SmartReader Plus 9 has the capability to count and record voltage
pulses and switch contact closure frequency. The SmartReader Plus 9 does not have an
internal temperature channel.
1.12.2
How the SmartReader Plus 9 Counts
The SmartReader Plus 9 has two channels. The channels can count the opening and closing
of external switch contacts and/or the occurrence of DC voltage pulses. Each channel has a
selectable slow or fast input. The SmartReader Plus 9 can only count pulses from one input at
a time on each channel.
The pulse logger can count up to 4095 switch contact closures or pulses over any sampling
interval. For example, if you select a sixteen second sampling interval, the SmartReader Plus
9 can record a maximum of 255 closures per second (4095/16). If the number of closures or
pulses exceeds 4095 during a sampling interval, the logger will record a count of 4095 for that
interval.
The logger keeps track of the number of times a switch opens and closes by continuously
maintaining a voltage potential across the input terminals. Every time the voltage changes
from high to low (i.e. switch closing) and back to high again (i.e. switch opening), the logger
increments its internal counter.
Counting voltage pulses (also called logic inputs) is similar to counting switch contacts, only