Choosing Input Channels - ACR Electronics SmartReader Plus Reference Manual

Table of Contents

Advertisement

Avoiding Ground Loops
Take special care to avoid ground loop problems when you use your SmartReader Plus 7. A
ground loop can occur when there is more than one path to ground in your logger-transducer
circuit. Ground loops can damage your SmartReader Plus 7 as well as your transducers. To
avoid ground loop problems in your SmartReader Plus 7 circuit:
1.
Do not use more than one grounded power supply to excite your transducers.
2.
Do not connect your logger to your computer for Realtime readings unless:
· your computer is battery-operated (i.e. not grounded) or
· your transducer power supply is not grounded.
If you must use more than one grounded power supply in your SmartReader Plus 7 circuit,
each transducer you use must be isolated. If you must use non-isolated transducers, then you
must use a Signal Repeater/Loop Isolator between the transducer output and the
SmartReader Plus 7 input.
Wiring Notes and Considerations
1.
The SmartReader Plus 7 has four common inputs (labeled "Common"). These are not
isolated from each other and can be used as the negative "-" connection for the power
supply (or battery) and any of the transducers being logged by the SmartReader Plus 7.
2.
The 200mV channels "float" when no connection is made to them. They can be expected
to read a positive voltage when left unconnected. When connected, however, they will
read the correct input voltage.
3.
The 25mA channels read positive current only. Make sure to observe polarity. If you are
using both channels simultaneously, ensure that both 4-20mA transmitters operate with
their negative (-) terminals tied together.
1.10.3

Choosing Input Channels

The SmartReader Plus 7 has seven external input channels to use in your particular
applications.
The resolution of your SmartReader Plus 7 logger is twelve bits. This means that it can
resolve analog signals into 4096 discrete steps. All channels have a resolution of 0.025% full
scale (1/4096 x 100%) which produces a very detailed graph in most applications.
The SmartReader Plus 7 can be used for measuring signals with a full scale output as low as
20mV on the 200mV channel (this translates to 1 part in 410 which is better than eight bit
resolution); this is useful for Wheatstone bridge type transducers which typically have a full
scale output in the tens of millivolts.
WARNING: The 200mV channel is more sensitive to noise signals generated by
the environment the logger is placed into - shielded cables should be used.
Copyright © 2010 ACR Systems Inc. All Rights Reserved.
SmartReader Plus Data Logger
52

Advertisement

Table of Contents
loading

Table of Contents