Smartreader Plus 3 Current, Voltage And Temperature Logger; Description - ACR Electronics SmartReader Plus Reference Manual

Table of Contents

Advertisement

SmartReader Plus Data Loggers
33
1.5

SmartReader Plus 3 Current, Voltage and Temperature Logger

The SmartReader Plus 3 is a versatile equipment performance logger. It monitors and
records AC current and voltage.
1.5.1

Description

The SmartReader Plus 3 has four current-monitoring channels, three 0-5 VDC inputs for
connection to AC voltage transducers, and one internal thermistor temperature channel. The
current-monitoring channels are used with external clamp-on-current probes to measure and
record alternating current.
The following current probes from Amprobe Instruments are compatible with the SmartReader
Plus 3:
· A60FL and A70FL for 60Hz power systems
· A65FL and A75FL for 50Hz operation
Use of these probes allows you to measure alternating current of conductors without the need
for breaking circuitry. For information on current ranges and resolutions possible with the
SmartReader Plus 3, refer to the
Current Channel Operation
The SmartReader Plus 3 measures electric current through external current probes. Each
current probe is a transformer with its internal coil serving as the secondary winding and the
current-carrying conductor being measured serving as the primary winding. The output of the
probe (which is the input to the SmartReader Plus 3) is conditioned through internal circuitry.
The probe produces an output of approximately 260 microamps full scale for each range
available on the current probe. Readings from the current probe are scaled into amperage
units through equations (in TrendReader software) that match the current probe and its range
setting.
Figure 3-1: SmartReader Plus 3
Specifications
in this chapter.