Delay After Backup - ACR Electronics SmartReader Plus Reference Manual

Table of Contents

Advertisement

To set this delay, click on the Delay Between Calls field and enter the value.
1.2.1.3.3

Delay After Backup

The Delay After Backup is the period of time that the alarm feature is temporarily disabled
after the logger has been backed up. The delay begins when you acknowledge the alarm by
backing up the logger that dialed out. The Delay After Backup deactivates the alarm, giving
you time to fix the problem that caused the alarm. The maximum delay is 720 minutes. When
you look at the logger status window (not in Edit Logger Setup) you will see the time counted
down. Note that the time shown may be slightly inaccurate if you Realtime the logger.
If the delay period elapses and the alarm condition still exists, the logger will dial out again. To
bypass the Delay After Backup period (if you have fixed the alarm problem but there is time
remaining on the delay which is preventing the alarm from being activated), you can clear the
logger to clear this delay.