Figure 3-2 4-Wire Resistance Test Connections - Keithley 2000-Scan Instruction Manual

Scanner card
Hide thumbs Also See for 2000-Scan:
Table of Contents

Advertisement

4-wire resistance tests
More precise measurements over a wider range of system and DUT conditions can be
obtained by using the 4-wire measurement scheme shown in
leads from the Model 2000 Multimeter are routed through the scanner to the resistor under test.
The extra set of sense leads minimizes the effects of voltage drops across the test leads, greatly
reducing measurement errors with lower DUT resistances. Note, however, that an extra two
poles of switching are required for each resistor tested. For this reason, only five resistors can be
tested using this configuration.
Figure 3-2
4-wire resistance
test connections
DMM Rear Panel
(Model 2001/2002 shown)
A. Test Configuration
HI
Sense HI
Sense LO
LO
2001
B. Simplified Equivalent Circuit
Figure
2000-SCAN
Out A
HI
1
1
Sense
5
5
HI
DUTs
(5)
LO
6
Sense
10
LO
Out B
R
DUT
2000-SCAN
Operation
3-17
3-2. Here, separate sense