Output Connections; Figure 2-3 Output Connections - Keithley 2000-Scan Instruction Manual

Scanner card
Hide thumbs Also See for 2000-Scan:
Table of Contents

Advertisement

Output connections

Use the supplied test leads for scanner output connections. Connect red leads to the output
(OUT A and OUT B) HI terminals, and connect black leads to the output LO terminals. See
Figure 2-3
leads" on page
it with the locking tab.
NOTE
After the scanner card is installed, the output leads must be connected to the multimeter rear
panel input jacks. See
Figure 2-3
Output connections
for details. Dress output test leads through the cable clamp, as described in
2-8. After all wires are connected and secure, close the plastic shield, and secure
If you intend to use the scanner card only in the 2-pole mode, it is not necessary to
connect output leads to both OUT A and OUT B. Use only OUT A for the 2-pole mode.
"Output connections to multimeter" on page 2-17
H L
H L
H L
H L
CH 1
CH 2
CH 3
CH 4
Card Connections and Installation
HI
To DMM
Input
LO
Red
Black
Red
H L
H L
H L
H L
H L
H L
CH 5
CH 6
OUT A
CH 7
CH 8
CH 9
CH 10
Note: OUT B connections not required
for 2-pole operation. Plastic shield
not shown.
"Dressing
for details.
HI
To DMM
Sense
LO
Black
H L
H L
OUT B
2-7

Advertisement

Table of Contents
loading

Table of Contents