Output Connections To Multimeter; Figure 2-11 2-Pole Output Connections - Keithley 2000-Scan Instruction Manual

Scanner card
Hide thumbs Also See for 2000-Scan:
Table of Contents

Advertisement

Output connections to multimeter

After installation, connect the scanner card output leads to the multimeter rear panel jacks as
follows:
WARNING
Figure 2-11
2-pole output
connections
NOTE
Signals applied to the front inputs of the multimeter cannot exceed ratings
of the scanner card as long as the scanner card output leads are connected
to the multimeter rear panel input jacks.
For 2-pole operation, connect OUT A HI (red) to INPUT HI, and connect OUT A LO
(black) to INPUT LO. See
Input HI
Out A
HI (Red)
Note: Card installation in a Model 2001/2002
is shown. For installation in Model 2000,
turn card over.
Be sure to select the rear inputs with the multimeter front panel INPUTS switch when
using the scanner.
Card Connections and Installation
Figure
2-11.
Input LO
Out A
LO (Black)
2-17

Advertisement

Table of Contents
loading

Table of Contents