Table 29. Sense command summary (cont.)
Command
:CURRent:AC
:AVERage
:ADVanced
[:STATe]?
[:STATe] <b>
[:STATe]?
:AUTO <b>
:AUTO ONCE
:AUTO?
:DETector
[:FUNCtion] <name>
[:FUNCtion]?
:CURRent:DC
:APERture <n>
:AUTO <b>
:AUTO ONCE
:AUTO?
:APERture?
:NPLCycles <n>
:AUTO <b>
:AUTO ONCE
:AUTO?
:NPLCycles?
:RANGe
[:UPPer] <n>
[:UPPer]?
:AUTO <b>
Description
Query state of advanced
filter.
Enable (1 or ON) or disable
(0 or OFF) filter.
Query state of digital filter.
Enable (1 or ON) or disable
(0 or OFF) auto filter.
Enable and then disable
auto filter.
Query auto filter.
Path to select function:
Select type of AC measure-
ment: RMS, AVERage.
Query detector function.
Path to configure DC current.
Specify integration rate in sec-
onds (166.67e-6 to 200e-3).
Enable (1 or ON) or disable
(0 or OFF) auto aperture.
Enable and then disable
auto aperture.
Query auto aperture (0 or 1).
Query aperture (integration
rate).
Specify integration rate (num-
ber of line cycles; 0.01 to
10).
Enable (1 or ON) or disable
(0 or OFF) auto NPLC.
Enable and then disable
auto NPLC.
Query auto line cycle inte-
gration (0 or 1).
Query line cycle integration
rate.
Path to configure measure-
ment range:
Select range (0 to 2.1).
Query range.
Enable (1 or ON) or disable
(0 or OFF) auto range.