Chapter 5 Temperature Calibration

to calibrate the temperature readings taken by the ATC probe. Since the temperature offset calculated during the
calibration is applied to all future temperature measurements, recalibrate if a different ATC probe is used.
Temperature calibration is for the calibration of an ATC probe (separate or built in to a pH electrode or ORP electrode.)
set the manual temperature value using the Manual Temp Value menu. The temperature calibration menu is not
accessible unless an ATC probe is connected to the meter.
2. Press
3. Press
traceable thermometers be used to measure and verify the temperature of the solution.
a. Press
b. Press f2 (done)
by the ATC probe on the ATC
readings on the Offset line. Press f2 (accept)
Thermo Scientific Orion Star A211 Benchtop and Star A221 and Star A321 Portable pH Meter Reference Guide
,
,
or
to highlight pH Channel and press f3 (select).
or
to highlight Temperature and press f3 (select).
or
to highlight Temperature Calibration and press f3 (select).
or
to highlight ATC
f3 (edit) to access the numeric entry screen and edit the temperature value.
,
,
or
to select the highlighted item and repeat until the temperature value read by the thermometers is
f1 (meas) to proceed to the measurement mode or press f2 (print) to proceed to the measurement mode
.
to highlight a number, decimal point or negative sign; press f3 (enter)
Reference Temp line, the original temperature read
f3 (select).