CSAT-5060 C-Band Transceiver
Ethernet-based Remote Product Management
AA.5.5.3.6 Config | LNA Page
Use this page to configure the Low Noise Amplifier (LNA) current alarm settings. At any time,
click Refresh to update this page and see the most recent data.
LNA Current Source
Use the drop-down list to select Enabled , Disabled or Offline LNA Disabled for the LNA Current
Source. Click Change to save the setting.
Current Window
Enter a percentage value from 20 to 50 in increments of 1 percent. If the monitored LNA Current
exceeds this value, an alarm occurs. If you set the value to 99, the LNA Current alarm monitoring
is disabled. Click Change to save the setting.
LNA Fault Logic
Use the drop-down list to select No Summary Fault or Summary Fault. Click Change to save the
setting.
Calibrate
Click Send the Command to set the calibration point for the LNA Current alarm.
Figure AA-11. Config | LNA page
AA–25
Revision 1
MN/CSAT5060.IOM