SECTION 12—BATTERIES
Batteries with terminal configuration as shown below MUST be used. Batteries that
have the reverse terminal configuration MUST NOT be used - otherwise injury and
damage may occur.
USE THIS CONFIGURATION
U1 Battery
POSITIVE (+) Battery Terminal
NOTE: Recommended battery type is spill proof and requires no maintenance except routine 
charging.
NOTE: Charge batteries daily. DO NOT allow batteries to completely discharge. To charge 
batteries, refer to Charging Batteries on page 62.
NOTE: Invacare recommends that both batteries be replaced if one battery is defective.

Removing/Installing the Batteries

NOTE: For this procedure, refer to FIGURE 12.1 on page 59.
Install protective caps on POSITIVE (+) and NEGATIVE (-) terminals.
All battery terminal covers (two on the right battery and two on the left battery)
MUST be installed prior to use.

Removing

1. Disconnect the joystick. Refer to Disconnecting/Connecting the Joystick on page 56.
2. Remove the seat. Refer to Removing/Installing the Seat Assembly on page 37.
3. Remove the shroud. Refer to Removing/Installing the Shroud on page 49.
4. Disconnect the front battery from the controller (BLACK connector). 
Pronto® M41 with SureStep®
WARNING
NEGATIVE (-)
U1 Battery
Battery Terminal
NEGATIVE (-)
Battery Terminal
WARNING
58
DO NOT USE
POSITIVE (+)
Battery Terminal
Part No. 1143206

Hide quick links: