9.2.1 Replacing the drive unit
6.01
6.02
9.2.2 Replacing the carbon brushes
•
Remove the seat and the rear shroud (see chapter 9.3.5).
•
Pull the de-clutching lever and jack up the vehicle.
•
Remove the screw that connects the release lever
to the drive unit.
•
Loosen the bolts that hold the wheels and
remove them, if necessary using a special tool to
remove wheels (see notes on page 24).
•
Release the clamps that hold the drive unit.
•
Disconnect the cable that connects the drive unit
and the electronic system, and remove the drive unit.
Reassembly is done in reverse order.
•
Remove the seat and the rear shroud (see chapter 9.3.5).
•
Remove the spray guard of the charger.
•
Remove all four screws on the charger
and remove the charger.
•
Remove both plastic screws from the drive unit
and replace the brushes.
Reassembly is done in reverse order.
Page 19
Service Manual