Auriga
Invacare
®
7
Error codes
Before assessing the error codes, carry out the following test:
•
Turn the remote on and off several times. Before you switch on wait approx. 5 seconds.
The test checks whether the error can be automatically rectifi ed by the electronics, and if necessary
deactivates the fl ashing status display on the control panel. If this is not the case, you can locate the fault
using the fl ash codes as follows:
•
Turn the control panel on and off.
•
Check which fl ash code is showing.
•
Wait for the fl ash code pause (approx. 2-3 seconds).
•
Re-count the code.
Flash
Fault
Code
1 x
Battery needs to be
charged
2 x
Battery voltage too
low
Seat lifter is raised
3 x
Battery voltage too
high.
4 x
Current time-out
5 x
Brake Fault
Impact on
Remarks
Scooter
No impact
• Battery charge is running low. Charge the
batteries as soon as possible.
Drive inhibited
• The batteries are depleted. Charge the batteries.
• If the scooter is left turned off for a time, then the
batteries may recuperate enough to allow it to be
driven for a very short period again.
Reduced
• Lower the seat lifter completely.
driving speed
Drive inhibited
• Battery voltage is too high. If a charger is
connected, disconnect it.
• The electronic system charges the batteries
when driving downhill and when braking. This
fault is caused by the battery being over-charged
because of this. Turn the scooter off and back
on again.
Drive inhibited
• The scooter has drawn too much current for too
long, possibly because the motor has been over-
worked, jammed or stalled. Turn the scooter
power off, leave it off for a few minutes, and then
turn the power back on again.
• The controller has detected a short-circuited
motor. Check the cable harness for short
and check the motor: Contact your authorised
Invacare® Dealer.
Drive inhibited
• Check that the de-clutching lever is in the
engaged position.
Page 13
Service Manual