SYSTEM TROUBLESHOOTING GUIDE
Product Water - Low Production Rate
Cause
Plugged pre-filters
Low water temperature
Low feed pressure
Fouled membrane
Membrane Troubleshooting Guide
The following chart illustrates the procedure for determination of RO membrane
performance. However, the chart represents only rough guidelines for determin-
ing performance of RO membrane. Depending on your tap water chemistry, the
rejection characteristics of the membrane may vary significantly.
Method of Calculate %
Testing
Rejection
TDS/
Measure feed water and
Conductivity RO product water TDS/
Tester
Conductivity
Alkalinity
Measure feed water and
Test Kit
RO product water
Alkalinity
Hardness
*Measure feed water
Test Kit
and RO product water
Hardness
*Caution: This test is not to be used on softened water sources.
Please feel free to contact us here at the factory for
any technical questions or problems you may have.
Toll-Free (also in Canada) 1-800-685-2783
SpectraPure®Inc .
E-mail: spectra@spectrapure.com Visit us on the web www.spectrapure.com
®
Corrective Action
Replace pre-filters
Heat feed water or use higher GPD membrane
Use booster pump or use higher GPD membrane
Replace membrane
Test Results
Is Rejection
greater than
95% ?
Is Rejection
greater than
90% ?
Is Rejection
greater than
90%?
Fax 480.894.6109 Fax us toll-free 1.877.527.7873
Conclusion
No - Replace Membrane
Yes - Membrane OK
No - Replace Membrane
Yes - Membrane OK
No - Replace Membrane
Yes - Membrane OK
17