SpectraPure
Membrane Output Calculation Example
What is the expected GPD from a 75 GPD System at 40 psi pressure and 60°F
water temperature?
PCF = 40 ÷ 60 = 0.666
TCF = 0.754 (from Table 1)
Expected GPD = 75 × 0.666 × 0.754 = 37.7 GPD ± 15%
Performance Test
The performance of a RO membrane is measured in terms of its rejection charac-
teristics.
Important: Test the quality of the membrane once every 6 months.
Note: This procedure will require a Conductivity Meter (TS-C61).
Procedure:
1.
Measure tap water conductivity. (Call it X)
2.
Run the system for 15-20 minutes.
3.
Rinse test instrument cell 2-3 times with RO water.
4.
Measure RO water conductivity. (Call it Y).
5.
Subtract RO water conductivity from tap water conductivity. (X - Y)
6.
Divide this quantity by tap water conductivity. (X - Y) ÷ X
7.
Rejection = [(X - Y) ÷ X ] ×100
* Conductivity in the above procedure could be replaced by hardness, alkalinity,
nitrate, phosphate, silica etc. (measured in ppm or mg/l).
Rejection of the RO Membrane Calculation Example
1.
Tap water hardness = 150 ppm (X)
2.
RO water hardness = 7 ppm (Y)
3.
X - Y = 143 ppm
4.
(X - Y) ÷ X = 143 ÷ 150 = 0.953
5.
Rejection = [ ( X - Y) ÷ X ] ×100 = 0.953 ×100 = 95.3
Membrane Hardness Rejection = 95.3 % : Membrane OK