PILLOW SPEAKER OPERATION

See page 24 for information on pillow speaker type selection.
See pages 17 and 25 to disable the TV's internal speakers or turn on pillow speaker sound.
The BM13II has a ¼" jack for pillow speaker connection. A one button analog pillow speaker, digital pillow speaker set for
Zenith or Philips code, HCI Visitor station or a Sip and Puff device can be used.
Pin Out
Sip and Puff control is via mechanical switch closure between Tip or Ring and Common.
Starting with the set off and switch closure between Ring and Common will turn the set on. Once set is on a momentary closure
will change the channel one channel. Maintaining the switch closure for about 3 seconds will change direction of the change.
For instance if channel change was up to the next higher channel the direction will change to down and change to the next lower
channel. The change is indicated on the screen. Holding the closure for about 5 seconds will turn the set off. A single button
pillow speaker will also function this way.
A switch closure between Tip and common will cause the volume to change. Maintaining the switch closure for about 3 seconds
will change direction of the change. For instance if volume change was up the direction will change to down. The change is
indicated on the screen. Holding the closure for about 5 seconds will mute the sound. The next closure will un-mute the sound.
A sip and puff device must be a two function device to control volume. Single function devices can only control channel change
and power.
NOTE: When the set is set for Sip and Puff operation audio is not available from the pillow speaker jack.

IR BANDS

There are four IR bands available. Bands one and two are for HCI IR remote controls. To change the IR band of the HCI IR
remote control press and hold the power button for 10 seconds. Bands 3 and 4 are for universal IR remotes. Band 3 is for
Magnavox codes and band 4 is for Apex codes. You can change IR bands is the setup menu (see page 24) or by pressing the
VOL+ and CH+ buttons on the front panel at the same time.