Heartway Medical Products PF2 User Manual page 23

Table of Contents

Advertisement

2
Battery
voltage too
low
3
Battery
voltage too
high
4
Current limit
time out
5
Brake fault
6
Out of
Neutral at
Power Up
7
Speed Pot
Error
8
Motor Volts
Error
9
Other Internal
Errors
Drive
Battery charge is empty. Recharge the batteries. If the
inhibited
scooter is left off for a few minutes, battery charge
may recover sufficiently to allow driving for a short
period of time.
Drive
Battery charge is too high. If a charger is plugged in,
inhibited
unplug it or turn the Charge/Run switch to Run.
Scooters powered by RHINO will charge the batteries
when traveling down slopes or decelerating. Excessive
charging in this manner may cause this fault. Turn the
scooter power off and then back on again.
Drive
The scooter has drawn too much current for too long,
inhibited
possibly because the motor has been over worked,
jammed or stalled. Turn the scooter power off, leave
for a few minutes, and then turn the power back on
again.
The controller has detected a shorted motor. Check the
loom for shorts and check the motor. Contact your
service agent.
Drive
Check that the park brake release lever is in the
inhibited
engaged position .
The park brake coil or wiring is faulty. Check the park
brake and wiring for open or short circuits. Contact
your service agent.
Drive
Throttle is not in neutral position when tuning switch
inhibited
key on. Return throttle to neutral, turn power off and
back on again. Throttle may need to be re-calibrated
Check throttle wiring.
Drive
The throttle or its wiring is faulty. Check for open or
inhibited
short circuits.
Throttle may not be correctly set up. Contact your
service agent.
Drive
The motor or its wiring is faulty. Check for open or
inhibited
short circuits.
Contact your service agent.
Drive
Contact your service agent.
inhibited
23