Summary of Contents for Omega HHM22

  • Page 1 OMEGA...
  • Page 2 European New Approach Directives. OMEGA will add the CE mark to every appropriate device upon certification. The information contained in this document is believed to be correct but OMEGA Engineering, Inc. accepts no liability for any errors it contains, and reserves the right to alter specifications without notice.
  • Page 3: Safety Information

    SAFETY INFORMATION The following safety information must be observed to insure maximum personal safety during the operation at this meter: 1. Do not use the meter if the meter or test leads look damaged, or if you suspect that the meter is not operating properly. 2.
  • Page 4: Specifications

    SPECIFICATIONS Display: 3½ digit liquid crystal display (LCD) with a maximum reading of 3200. Analog bar graph: 32 segments with measurements 12 times per second. Polarity: Automatic, (-) negative polarity indication. Overrange: "OL" mark indication. Low battery indication: The " "...
  • Page 5 DC VOLTS Ranges: 320mV,3.2V,32V,320V,600V Resolution: 100µV Accuracy: ±(0.25%rdg + 1dgt) Input impedance: >10MW Overload protection: 600VDC or AC rms AC VOLTS (50Hz - 400Hz) Ranges: 3.2V,32V,320V,600V (3.2V only @50Hz-60Hz) Resolution: 1mV Accuracy: ±(1.0%rdg + 4dgts) on 3.2V to 320V ranges ±(2.0%rdg + 4dgts) on 600V range Input impedance: >10MW Overload protection: 600VDC or AC rms...
  • Page 6 AC CURRENT (50Hz - 400Hz) Ranges:320µA, 32mA, 10A Accuracy: ±(1.5%rdg + 4dgts) on µA,mA ranges ±(3.5%rdg + 4dgts) on 10A range Input protection: 0.5A / 250V fast blow ceramic fuse 10A / 600V fast blow ceramic fuse RESISTANCE Ranges: 320W ,3.2KW ,32KW ,320KW ,3.2MW ,32MW Accuracy: ±(0.3%rdg + 3dgts) on 320W range ±(0.3%rdg + 1dgt) on 3.2KW to 320KW ranges ±(1.2%rdg + 1dgt) on 3.2MW range...
  • Page 7 CAPACITANCE Ranges: 3.2nF,32nF,320nF,3.2µF,32µF Accuracy: ±(3.0%rdg + 10dgts) Test volts: 50mV/50Hz TRANSISTOR hFE Ranges: 0 - 1000 Base current: 10µAdc approx. (Vce=3.0Vdc)
  • Page 8: Operation

    OPERATION Before taking any measurements, read the Safety Information Section. Always examine the instrument for damage, contamination (excessive dirt, grease, etc.) and defects. Examine the test leads for cracked or frayed insulation. If any abnormal conditions exist do not attempt to make any measurements.
  • Page 9 Voltage Measurements 1. Connect the red test lead to the "VW " jack and the black test lead to the "COM" jack. 2. Set the Function switch to the desired voltage type (DCV) or (ACV) position. 3. Touch the probes to the test points, the range will change automatically to the level that will display the input voltage with best resolution.
  • Page 10 Resistance Measurements 1. Set the Function switch to the " W " position. 2. Turn off power to the circuit under test. External voltage across the components causes invalid readings. 3. Connect the red test lead to the "VW " jack and the black test lead to the "COM"...
  • Page 11 Continuity Measurements 1. Set the Function switch to " " position. 2. Turn off power to the circuit under test. External voltage across the components causes invalid readings. 3. To toggle between the continity/diode modes, press Range Switch. 4. Connect the test leads to the two points at which continuity is to be tested. The buzzer will sound if the resistance is less than approximately 20W ±5W .
  • Page 12: Maintenance

    MAINTENANCE WARNING Remove test leads before changing battery or fuse or performing any servicing. Battery Replacement Power is supplied by a 9 volt "transistor" battery. (NEDA 1604, IEC 6F22). The " " appears on the LCD display when replacement is needed. To replace the battery, remove the two screws from the back of the meter and lift off the battery case.