How To Measure Blood Pressure; Single Mode; Avg. Mode; Manu. Mode - Omron HEM-907 Instruction Manual

Digital automatic blood pressure monitor
Hide thumbs Also See for HEM-907:
Table of Contents

Advertisement

If any abnormal matter occurs during the mea-
surement such as the inÀ ation does not stop,
remove the arm cuff or pull the air tube from the
main unit.
You may suffer peripheral neuropathy.
Requests from OMRON
Be sure that the patient should not touch the unit.
Be sure to monitor all the time that the unit and the
patient are not abnormal.
Do not use the unit in a vehicle.
List of Measurement Modes
To measure
only once
To measure three
times
(or two times)
consecutively
To measure
by using a
stethoscope
To con¿ rm the
accuracy of
pressure display
Warning

How to Measure Blood Pressure

In the following case, con¿ rm the measurement
with the stethoscope.
(1) When an irregular pulse wave is displayed
· An error may be displayed when an external vibra-
tion is applied or the patient moves his/her body.
(2) When an error is generated or the measured value
is doubtful
· An error may be displayed when an external vibra-
tion is applied or the patient moves his/her body.
· An error may be generated in the measured valueor
the blood pressure may not be able to measure for
the patient having peripheral circulatory disturbance
or extreme hypotension.
· An error may be generated in the measured valueor
the blood pressure may not be able to measure for
the patient having short variation of blood pressure
such as arrhythmia.
Do not use a portable phone near the unit.
The unit may malfunction.

SINGLE Mode

AVG. Mode

MANU. Mode

CHECK Mode

Caution
Refer to Page
16.
Refer to Page
17.
Refer to Page
18.
Refer to Page
19.
15