3.5 CONNECTING MAIN POWER
Connect the ac main power connections as shown in Figure 3-8.
Warning: Do not connect AC power to your device until you have
completed all input and output connections. This device must only be
installed by a specially trained electrician with corresponding qualifications.
Failure to follow all instructions and warnings may result in injury!
Figure 3-8 Main Power Connections (ac powered unit)
Table 3-2 shows the wire color and respective terminal connections for both USA and
Europe.
Table 3-2 Main Power Connections (ac powered unit)
TB1
7
8
9
AC POWER
~ ac Line
~ ac Neutral
~ ac Earth
WIRE COLORS
EUROPE
Brown
Blue
Green/Yellow
17
USA
Black
White
Green