After a beep, you will see the
meter count down 5 seconds
until the test is complete and
your control test result is
displayed.
The meter will automatically
recognize and
mark the
control result for you.
NOTE:
The
does not
indicate the control test
result is within the speci-
fied range. It only marks it
as a control solution test.
It will not be included in
your averages.
Compare your control test
result with the Normal Control
Range printed on the test
strip bottle label or on the
bottom of the test strip box.
NOTE:
Low and High
control range values can
be found on the bottom of
the test strip box.
16
Need Help? Call toll free: 1-800-348-8100
If your control test result is out of range, do not use
your meter for blood glucose testing until you resolve
the issue. See "What You Should Do" on page 36.
To turn your meter off, simply
remove the test strip. The
result will be automatically
marked and stored in the
meter memory. Control results
will not be included in your
blood glucose averages or in
the 7 day HI-LO summary.
Carefully dispose of the used
test strip.
IMPORTANT: To ensure accuracy of your blood glucose
test results and accuracy of your control test results,
use only C
®
control solutions.