Do not overload wall outlets and extension cords as this may result in
fire or electric shock.
Mains lead protection (U.S.: Power cord): Power cords should be routed
so that they are not likely to be walked upon or pinched by items
placed upon or against them, paying particular attention to cords at
plugs and receptacles.
Use a power cord that matches the voltage of the power outlet, which
has been approved and complies with the safety standard of your
particular country.
External equipment
External equipment intended for connection to signal input/output or
other connectors, shall comply with relevant UL/ EN/ IEC standard (e.g.
UL/EN/IEC 60950 for IT equipment and ANSI/AAMI ES60601-1/EN
60601-1 / IEC 60601 series for medical electrical equipment). In
addition, all such combinations – systems – shall comply with the
standard IEC 60601-1-1, Safety requirements for medical electrical
systems. Equipment not complying with ANSI/AAMI ES60601-1/EN /
IEC 60601-1 shall be kept outside the patient environment, as defined
in the standard.
Equipment not complying with IEC 60601 must be kept outside the
patient environment, as defined in the standard as at least 1.5 meters
from the patient or the patient support.
Any person who connects external equipment to signal input, signal
output, or other connectors has formed a system and is therefore
responsible for the system to comply with the requirements of IEC
60601-1-1. If in doubt, speak with a qualified technician.
In locations where 240V outlets are used, connect this display only on a
center-tapped, 240V, single-phase supply.
9