Degree of Safety in the Presence of Flammable Anesthetic Mixture with
Air or with Oxygen or Nitrous Oxide: Not suitable for use in the
presence of a flammable anesthetic mixture with air or with oxygen or
nitrous oxide.
Mode of Operation: Continuous.
Intended use
This display is an AMLCD display designed for viewing medical X-ray
images. This unit should not be used near patients and should be kept
outside of 1.83 m perimeter and 2.29 m vertical.
The unit is not intended to be used in any life support system and
diagnosis.
Power connection
Power requirements: The display must be powered using a medical
approved 12VDC power supply (MDRC-1119, MDRC-1119TS,
MDRC-1119HB) or a 24VDC power supply (MDRC-2120).
The medical approved DC power supply must be powered by the AC
mains voltage.
Power requirements: The display must be powered using a medical
approved 24 VDC SELV power supply.
The medical approved DC power supply must be powered by the AC
mains voltage.
Power cords
Utilize a UL-listed detachable power cord, 3-wire, type SJ or equivalent,
18 AWG min., rated 300 V min., provided with a hospital-grade type
plug 5-15P configuration for 120V application, or 6-15P for 240V
application.