Calibration Mode

Calibration (CAL) Mode
Warning! Entering into
this mode can erase
the calibration already
saved. You need
approved calibration
weights to use
calibration mode.
Note: If this procedure
is attempted without
proper calibration
weights applied, the
scale will abort the
process and retain the
original calibration
data.
10. Press the UNITS key.
Parity selection is stored.
11. Press the ZERO key...
PROT is displayed.
12. Press the UNITS key...
The current setting is displayed. Use the
ZERO key to toggle between the four
choices: NCI STD, NCI SMA, AS350D,
PS6L.
13. Press the UNITS key...
Protocol selection is stored.
14. When finished configuring your scale, press
the ZERO key until the display shows DONE,
then press the UNITS key.
Or, move Switch 1 to CLOSED position for
normal weighing mode.
The calibration (CAL) mode menu lets you
calibrate your scale. The items in the calibration
menu are as follows:
Pounds/Kilograms (LB or 1000g)
Selects the unit of measure of your calibra-
tion test weights.
Scale or Classifier
When calibrating the scale for LB, you are
able to calibrate the unit as a scale or as a
classifier (weight classifier).