Channel Settings - The General Tab; Output Impedance; Trigger Delay; Amplitude Correction Factor - LeCroy ArbStudio Manual

Arbitrary waveform generator
Table of Contents

Advertisement

ArbStudio 
The Settings screen is divided into the following tabs: 
1. The General Tab (below) 
2. The Run Mode Tab (Waveform Sequence) (on page 34) 
3. The Run Mode Tab (Modulation) (on page 36) 
4. The Trigger IN Tab (on page 38) 
5. The Trigger OUT Tab (on page 39) 

Channel Settings - The General Tab

The General tab is shown by default when the Channel Settings screen is opened. 
Sections of the Channel Settings General tab are explained as follows: 

Output Impedance

It allows setting the channel output impedance and it can be set at: 50 Ohm, Low or High Impedance.  

Trigger Delay

You can set a delay from the trigger event (start) for each analog output channel. 

Amplitude Correction Factor

The Amplitude Correction Factor (CF) consider the deviation of the load and of the channel source resistance 
from the nominal 50 Ohm resistance. 
Follow the next steps to evaluate the CF: 
• Apply to the load a fixed 5 V nominal voltage using the strobe functionality of the selected channel 
• Read the voltage load on the multimeter 
 