Range Specifications - Extech Instruments EX310 User Manual

Mini multimeter with non-contact voltage detector (ncv)
Hide thumbs Also See for EX310:
Table of Contents

Advertisement

Range Specifications

Function
Range
Resolution
Non-contact
100 to 600V Resolution & accuracy do not apply since the
AC Voltage
meter does not display the voltage in this mode.
detector
The lamp at the top of the meter's display flashes
when voltage is sensed and an audible warning will
sound
DC Voltage
200mV
0.1mV
(V DC)
2000mV
1mV
20V
0.01V
200V
0.1V
600V
1V
AC Voltage
200V
0.1V
(VAC)
600V
1V
50/60Hz
DC Current
200mA
0.1mA
(A DC)
10A
0.01A
AC Current
200mA
0.1mA
(A
AC)
10A
0.01A
50/60Hz
Resistance
200Ω
0.1Ω
2000Ω
20kΩ
0.01kΩ
200kΩ
0.1kΩ
2000kΩ
1kΩ
Notes:
Accuracy specifications consist of two elements:
• (% reading) – This is the accuracy of the measurement circuit.
• (+ digits) – This is the accuracy of the analog to digital converter.
o
o
Accuracy is stated at 65
F to 83
F (18
Accuracy
±(0.5% reading + 2 digits)
±(1.0% reading + 2 digits)
±(1.5% reading + 2 digits)
±(1.5% reading + 3 digits)
±(2.0% reading + 4 digits
±(1.5% reading + 2 digits)
±(2.5% reading + 5 digits)
±(1.8% reading + 5 digits)
±(3.0% reading + 7 digits)
±(1.2% reading + 4 digits)
±(1.2% reading + 2 digits)
o
o
C to 28
C) and less than 75% RH.
14