Extech Instruments EX360 User Manual

Extech Instruments EX360 User Manual

True rms digital multimeters ex360 series

Advertisement

USER MANUAL 
True RMS Digital Multimeters 
EX360 Series 
 
EX360 True RMS Digital Multimeter 
EX363 True RMS DMM with Temperature and µA AC/DC 
EX365 True RMS DMM with 10A AC/DC Current 
 
 
 
Additional User Manual Translations available at www.extech.com 

Advertisement

Table of Contents
loading

Summary of Contents for Extech Instruments EX360

  • Page 1 USER MANUAL  True RMS Digital Multimeters  EX360 Series    EX360 True RMS Digital Multimeter  EX363 True RMS DMM with Temperature and µA AC/DC  EX365 True RMS DMM with 10A AC/DC Current        Additional User Manual Translations available at www.extech.com ...
  • Page 2: Table Of Contents

    Table of Contents  1.  INTRODUCTION  3  2.  SAFETY INFORMATION  4  3.  DESCRIPTIONS  6  4.  OPERATION  9    Powering the Meter    Disable Auto Power OFF    Display Backlight    Smart Data Hold    Test Lead Considerations    Voltage Measurements    Lo Z Voltage Measurements    PEAK MAX‐MIN HOLD Mode (AC only)    10A AC/DC Current Measurements (EX365 only)    µA AC/DC Current Measurements (EX363 only)    Non‐Contact Voltage Detector    Resistance Measurements    Continuity Measurements    Capacitance Measurements ...
  • Page 3: Introduction

    1.  Introduction  Thank you for selecting the Extech EX360 Series Meter.      The EX360 Series are feature‐packed True RMS digital multimeters. In addition to standard DMM  features, they offer a Low Impedance (Lo Z) mode, Smart Data Hold, Automatic AC/DC voltage  Detect mode, Backlit LCD, and a Non‐Contact Voltage Detector that senses electrical sources  safely.     This device is shipped fully tested and calibrated and, with proper use, will provide years of  reliable service. Please visit our website (www.extech.com) to check for the latest version of this  User Guide, Product Updates, Product Registration, and Customer Support.    Features   6000 count digital display and 60 segment analog bar graph display   Large backlit LED display   True RMS ACV measurements   Lo Z mode (with Auto AC/DC Voltage detect) protects readings from ghost voltages   Auto and Manual Range modes   0.5% DCV accuracy   Smart Data Hold   Peak Hold   Relative mode   Auto Power OFF (APO) with disable function   Temperature measurements (Model EX363 only) with included temperature probe   AC/DC µA function (Model EX363 only)   10A AC/DC current measurements with 11A/1000V high energy fuse (EX365 only) ...
  • Page 4: Safety Information

    2. Safety Information  To ensure the safe operation and service of the meter, follow these instructions closely. Failure to  observe warnings can result in severe injury.    WARNINGS  WARNINGS identify hazardous conditions and actions that could cause BODILY HARM or DEATH.   When handling test leads or probes, keep hands and fingers behind the finger guards at all times.   Remove test leads from the meter before opening the battery compartment or meter housing.   Use the meter only as specified in this User Guide or accompanying Quick Start to avoid compromising  the protections provided by the meter.   Be sure to use the proper terminals, switch positions, and ranges when taking measurements.   Verify the meter’s operation by measuring a known voltage. Have the meter serviced if the meter  responds unusually or if there are questions regarding the meter’s functional integrity.   Do not apply more than the rated voltage, as marked on the meter, between terminals or between any  terminal and earth ground.   Replace blown fuses with fuses of the same type and rating as specified in this User Guide.   Use caution working with voltages above 30 VAC RMS, 42 VAC peak, or 60 VDC. These voltages pose a  shock hazard.   To avoid misleading readings that could lead to electric shock and injury, replace the batteries as soon as  the low battery indicator is displayed.   Disconnect power to the circuit under test and discharge all high‐voltage capacitors before testing  resistance, continuity, diodes, or capacitance.   Do not use the meter in the presence of explosive gas or vapor.   To reduce risk of fire or electric shock, do not use the meter if it is wet and do not expose the meter to  moisture.   Individual protective equipment should be used if HAZARDOUS LIVE parts in the installation where  measurements are to be carried out could be accessible. ...
  • Page 5 Safety Symbols that are typically marked on meters and instructions  This symbol, adjacent to another symbol, indicates the user must refer to the    manual or user guide for further information.  Risk of electrical shock     Fuse symbol      Equipment protected by double or reinforced insulation  Battery symbol    Conforms to EU directives    Do not discard this product in household trash.    AC measurement    DC measurement  Earth ground      Unsafe Voltage Alert  When the meter detects a voltage equal to or greater than 30V or a voltage overload (OL) in V, mV, or Lo Z  mode, the symbol  is displayed. This system was designed to alert the user of a potentially hazardous  voltage.  PER IEC1010 OVERVOLTAGE INSTALLATION CATEGORY  OVERVOLTAGE CATEGORY I  Equipment of OVERVOLT AGE CATEGORY I is equipment for connection to circuits in which measures are  taken to limit the transient over‐voltages to an appropriate low level.  Note – Examples include protected electronic circuits.  OVERVOLTAGE CATEGORY II  Equipment of OVERVOLTAGE CATEGORY II is energy‐consuming equipment to be supplied from the fixed  installation.  Note – Examples include household, office, and laboratory appliances.  OVERVOLTAGE CATEGORY III  Equipment of OVERVOLTAGE CATEGORY III is equipment in fixed installations.  Note – Examples include switches in the fixed installation and some equipment for industrial use with  permanent connection to the fixed installation.  OVERVOLTAGE CATEGORY IV ...
  • Page 6: Descriptions

    3. Descriptions    Meter Description (EX365 pictured)  Non‐Contact Voltage Detector and Continuity alert light  LCD multi‐function display  M (MODE) button (also  F button on EX363)  R RANGE and Lo/Hi button  Rotary function select switch  Positive input terminal (10A Current) for EX365 only  Common (‐) input terminal  Positive input terminal: Voltage, Resistance, Capacitance, Temperature (EX363), and µA  (EX363)  LCD Backlight button  H Data Hold button  PEAK MAX‐MIN /  REL (RELATIVE) button  Battery compartment on back of meter    Fig 3‐1 METER DESCRIPTION    EX36x‐en‐GB_V1.6   1/17 ...
  • Page 7 Display Icon Descriptions   Auto Volts: Automatic AC/DC voltage detect (for Lo Z mode only)      Auto: Automatic range mode  HOLD: Display hold   Lo Z: Low Impedance mode   : Relative mode    and AC: Alternating Current Measurements    and DC: Direct Current Measurements    : Main display digits    : Bar graph   A: Amperes (Current)   V: Volts (Voltage)   F: Temperature units   F: Farads (Capacitance measurement units)   Hz: Hertz (Frequency measurements unit)   Ω (Ohms): Resistance measurement unit   :Battery status icon   ‐ Minus (negative) sign  ...
  • Page 8 Push‐Button Descriptions    Press the M (MODE) button to perform the following. The F function is only available on the EX363 model: Switch Position  MODE (M) Button Function  V   Hz      A   Hz    (EX365)     Ω Ω             ...
  • Page 9: Operation

    4. Operation  CAUTION: Read and understand all of the Safety statements listed in the safety section  of this manual prior to use.  Powering the Meter  Turn the rotary function switch to any position to power the meter. Check the batteries if the  unit fails to power ON. Refer to the Maintenance section for battery and fuse replacements.  Turn the function switch to the OFF position to power OFF the meter.  The meter has an Auto Power OFF feature (APO) where the meter switches OFF after 20  minutes of inactivity. When APO is enabled, the APO icon will show on the display when the  meter is powered ON. To disable APO, refer to the next section.  Note: The meter displays the battery capacity when powering up. The display will either show the  word FULL (battery fully charged) or a percentage representing remaining battery power. The  international segmented battery strength symbol  is also shown on the display at all times when  the meter is ON and when there is sufficient battery power to power the meter.  Disable Auto Power OFF  The meter will auto power OFF after 20 minutes of inactivity. To defeat this feature, follow the  steps below.  With the meter OFF, press and hold the M (MODE) button and, while continuing to hold, turn  the rotary function switch to any position to power ON the meter.  The display will show AOFF.  Release the button.  APO will now be disabled until the next cycle of power.    Display Backlight  With the meter powered ON, press the backlight button  to switch the backlight ON or OFF. Note  that excessive use of the backlight will shorten the battery life.   Smart Data Hold  To freeze the LCD meter reading, press the H (HOLD) button. While data hold is active, the HOLD  display icon appears on the LCD. Press the H button to return to normal operation. The HOLD icon  will switch OFF.  The meter’s beeper will sound and the LCD will flash if the detected signal is 50 counts greater than  the held reading.  Test Lead Considerations  Test lead probe covers can be removed for CAT II 1000V installations. Use the test lead probe  covers for CAT III 1000V or CAT IV 600V installations. Do not measure voltages > 1000V AC or DC.  Remove the storage caps from the meter end of test leads before connecting leads to the meter. EX36x‐en‐GB_V1.6   1/17 ...
  • Page 10: Voltage Measurements

    Voltage Measurements  WARNING: Remove the test lead probe covers for CAT II 1000V installations. Use the test  lead probe covers for CAT III 1000V or CAT IV 600V installations. Do not measure voltages  greater than 1000V.  CAUTION: When connecting the test leads to the circuit or device under test, connect the  black lead before the red; when removing the test leads, remove the red before the black lead.  AC Voltage Measurements  1. Insert the black test lead banana plug into the negative (COM) jack and the red test lead  banana plug into the positive (V/Ω) jack.  2. Turn the function switch to the   position.  3. The meter defaults to the Auto Range mode (Auto displayed on LCD). Press the R button to  access the manual range mode. Press and hold the R button to return to the Auto Range  mode (Auto will be displayed).  4. Read the Warning and Caution statements above to determine whether or not to use the  test lead probe covers.  5. Touch the test probe tips to the circuit under test.   6. Read the digital value and the bar graph representation of the measurement in the display.  The display will also indicate the proper decimal point and measurement type/units symbols.    7. Note the voltage alert symbol   when voltage is present.  8. Use the M button to view the frequency (Hz) of the measured voltage.    Fig 4‐1 AC VOLTAGE MEASUREMENTS   EX36x‐en‐GB_V1.6   1/17 ...
  • Page 11 DC Voltage Measurements    Insert the black test lead banana plug into the negative (COM) jack and the red test lead  banana plug into the positive (V/Ω) jack.  Move the Function Switch to the  or   position.   Read the Warning and Caution statements at the beginning of this section to determine  whether or not to use the test lead probe covers.  Touch the test probe tips to the circuit under test. Be sure to observe the correct polarity  (red lead to positive, black lead to negative).  The meter defaults to the Auto Range mode (Auto displayed on LCD). Press the R button  to access the manual range mode. Press and hold the R button to return to the Auto  Range mode.  Read the digital value and the bar graph representation of the measurement in the  display. The display will also indicate the proper decimal point and measurement  type/units symbols. If the polarity is reversed, the display will show (‐) minus before the  value.    Note the voltage alert symbol   when voltage is present.  The meter is capable of detecting DC voltages to 1000V.    Fig 4‐2 DC VOLTAGE MEASUREMENTS     EX36x‐en‐GB_V1.6   1/17 ...
  • Page 12: Lo Z Voltage Measurements

    Lo Z Voltage Measurements  When the function switch is turned to the Lo Z   position, the meter incorporates an  automatic voltage detection circuit to automatically determine voltage AC or DC and a low Z  (impedance) circuit* that eliminates ghost voltage complications. Refer to the Voltage  Measurements section earlier in this guide for Safety information and connection diagrams.  *The Lo Z impedance is approx. 3kΩ increasing to over 100kΩ when measuring 1000V.  Insert the black test lead banana plug into the negative (COM) jack and the red test lead  banana plug into the positive (V/Ω) jack.  Move the Function Switch to the Lo Z   position.   Touch the test probe tips to the circuit under test. Be sure to observe the correct polarity  (red lead to positive, black lead to negative).  The meter defaults to the Auto Range mode (Auto displayed on LCD). Manual range  cannot be used in this mode.   Read the digital value and the bar graph representation of the measurement in the  display. The display will also indicate the proper decimal point and measurement  type/units symbols. If the polarity is reversed, the display will show (‐) minus before the  value.  The meter is capable of detecting AC/DC voltages to 1000V.    Note the voltage alert symbol   when voltage is present.    PEAK MAX‐MIN HOLD Mode (AC only)  In the PEAK HOLD mode the meter records MAX and MIN measurement values. When the  measurement exceeds a recorded MAX or MIN value, the meter replaces the existing value  with the newer value.  1. Press and hold the PEAK button for at least two seconds. PEAK MAX MIN will be displayed  and the meter will begin recording MAX and MIN values.  2. A short press of the PEAK button will show the MAX recorded reading.  3. Another short press of the PEAK button will show the MIN recorded reading.  4. Press the PEAK button again to view real time readings while MAX MIN recording  continues in the background.  5. To pause MAX MIN recording, press the H button. The display will freeze and the HOLD  icon will appear. Press H again to return to real time readings (with MAX MIN recording ...
  • Page 13: A Ac/Dc Current Measurements (Ex365 Only)

    10A AC/DC Current Measurements (EX365 only)  WARNING: Do not handle the test leads above the finger/hand guard barrier.  CAUTION: Observe CAT III 1000V CAT IV 600V with respect to Earth Ground.  Insert the black test lead into the COM terminal and the red test lead into the A terminal.  Turn the meter’s function switch to the   or    position. The A symbol will appear on the  display indicating Amperes (Amps).  The meter defaults to the Auto Range mode. When in Auto mode, the display shows the Auto  icon in the upper left corner. Use the R button to manually range the meter. Press and hold  the R button to return to the Auto Range mode.  The meter display will show AC or DC depending on the function switch position.  Current measurements must be taken in series with the circuit under test. See accompanying  diagram.  Touch the black test lead to the negative side of the circuit and the red test lead to the  positive side of the circuit.  Read the current measurement in the display represented by numerical digits and bar graph.  The display will indicate the proper decimal point and value. In DC, If the polarity is reversed,  the display will show (‐) minus before the value.   Use the M button to view the frequency (Hz) of the measured current.    Fig 4‐3 AC/DC 10A CURRENT MEASUREMENTS  EX36x‐en‐GB_V1.6   1/17 ...
  • Page 14: Μa Ac/Dc Current Measurements (Ex363 Only)

    µA AC/DC Current Measurements (EX363 only)  WARNING: Do not handle the test leads above the finger/hand guard barrier.  CAUTION: Observe CAT III 1000V and CAT IV 600V with respect to Earth Ground.  Insert the black test lead into the COM terminal and the red test lead into the µA terminal.  Turn the meter’s function switch to the µA position. The µA units symbol will appear on the  display indicating that micro‐amperes are being measured.  The meter defaults to the Manual Range mode. Auto range cannot be used in this mode.   The meter display will show AC or DC depending on the function switch position.  Current measurements must be taken in series with the circuit under test. See accompanying  diagram.  Touch the black test lead to the negative side of the circuit and the red test lead to the  positive side of the circuit.  Read the current measurement in the display represented by the numerical digits and the bar  graph. The display will indicate the proper decimal point and value. In DC, If the polarity is  reversed, the display will show (‐) minus before the value.    Fig 4‐4 µA AC/DC CURRENT MEASUREMENTS  EX36x‐en‐GB_V1.6   1/17 ...
  • Page 15: Non-Contact Voltage Detector

    Non‐Contact Voltage Detector  WARNING: It is possible for voltage to be present in a circuit even if the meter does not  beep or light the NCV LED lamp at the top of the meter. Always verify meter operation on a  known live AC current circuit and verify that the batteries are fresh before use.  The  audible  beeper  sounds  and  the  LED  lamp  at  the  top  of  the  meter  lights  when  the  meter  senses an electrical voltage field.  If the meter does not emit a tone or light the LED in this mode, there is still the possibility that  voltage is present. Please use caution.  Turn  the  function  switch  to  the  NCV  position  to  access  the  Non‐Contact  Voltage  Detect  mode.  Use ...
  • Page 16: Resistance Measurements

    Resistance Measurements  Cautions: Switch OFF power to the device under test before measuring. Do not test on circuits  or devices where 60VDC or 30VAC is present.  Insert the black test lead banana plug into the negative (COM) jack. Insert the red test  lead banana plug into the positive (V/Ω) jack.  Turn the Function Switch to the Ω position.  Use the M button to select the Ω icon on the display indicating resistance only (without  the audible continuity icon showing).   Touch the test probe tips across the circuit or part under test. It is best to disconnect one  side of the part under test so the rest of the circuit will not interfere with the resistance  reading.  Read the resistance value in the display. The display will indicate the proper decimal  point and value. If the reading is out of range, the OL display icon will appear.    Fig 4‐6 RESISTANCE/CONTINUITY MEASUREMENTS     Continuity Measurements   Insert the black test lead into the negative COM terminal and the red test lead into the  positive terminal.  Set the function switch to the   position.  Use the M button to select the Continuity mode. Look for the Continuity icon   on the  display.  Touch the test probe tips across the wire or circuit under test.  If the resistance is < 20 Ω, the beeper will sound and the LED lamp at the top of the meter  will flash. The beeper and LED lamp will switch off if the measured signal rises above  approximately 200 Ω. For an open circuit condition the meter will display  .  EX36x‐en‐GB_V1.6   1/17 ...
  • Page 17: Capacitance Measurements

    Capacitance Measurements  WARNING: To avoid electric shock, remove power to the circuit under test and discharge the  capacitor under test before measuring.  Do not test on circuits or devices where 60VDC or  30VAC is present.  Set the function switch to the  capacitance position.  Insert the black test lead banana plug into the negative COM jack   and the red test lead banana plug into the positive   jack.  Press the M button to select the unit of measure symbol F.  Touch the test probe tips across the part under test.  Use the Relative mode   to compare measurements to a saved, known capacitance value.  See next section for Relative mode instructions.  Read the capacitance value in the display.  The display will indicate the proper decimal point and value.  Note: For large values of capacitance several seconds may be required before the reading  stabilizes.    Fig 4‐7 CAPACITANCE MEASUREMENTS   Relative Mode  In Relative mode a reference reading can be stored with which to compare subsequent readings.  Press the   button to store the displayed reading in memory, this becomes the reference. The  symbol  will be displayed when the Relative mode is active. Subsequent readings will now be  compared to the stored reference (Displayed Reading = Measurement minus Reference). Press the   button again to exit the Relative mode; the Relative symbol will switch off.    EX36x‐en‐GB_V1.6   1/17 ...
  • Page 18: Diode Test

    Diode Test  Insert the black test lead banana plug into the negative COM jack and the red test   lead banana plug into the   positive jack.  Turn the function switch to   position. Use the M button to select the diode function if  necessary (the diode and voltage symbols will appear on the LCD when in Diode test mode).  Touch the test probe tips to the diode or semiconductor junction under test. Note the  meter reading.  Reverse the test lead polarity by reversing the red and black leads. Note this reading.  The diode or junction can be evaluated as follows:   If one reading displays a value (typically 0.400V to 0.900V) and reverse reading displays  OL, the diode is good.   If both readings display OL the device is open.   If both readings are very small or ‘0’, the device is shorted.    Fig 4‐8 DIODE TESTING       EX36x‐en‐GB_V1.6   1/17 ...
  • Page 19: Temperature Measurements (Ex363 Only)

    Temperature Measurements (EX363 only)  Insert the supplied temperature probed into the COM and positive terminals observing  correct polarity.  Turn the function switch to the temperature position. Use the  F button to select the  desired unit of measure.    Touch the temperature probe tip to the device under test or leave the temperature probe  in the open air to measure ambient temperature.  Read the temperature measurement on the LCD.  To set the currently selected unit of measure as default, press and hold the backlight button  while turning the meter ON.  TEMP   Fig 4‐9 TEMPERATURE MEASUREMENTS     EX36x‐en‐GB_V1.6   1/17 ...
  • Page 20: Maintenance

    5. Maintenance   WARNING: To avoid electrical shock, remove the test leads, disconnect the meter from  any circuit and turn OFF the meter before opening the case. Do not operate with an open case.  Battery Replacement  Remove the test leads from the meter.  Remove the meter’s protective cover.  Remove the two Phillips head screws that  secure the battery compartment at the  back of the meter.   Open the battery compartment and replace  the 9V battery observing correct polarity.  Re‐assemble the meter before use  Safety: Please dispose of batteries responsibly;  never dispose of batteries in a fire, batteries  may explode or leak. If the meter is not to be  used for 60 days or more, remove the battery  and store separately.                Fig 5‐1 BATTERY REPLACEMENT        Never dispose of used batteries or rechargeable batteries in household waste.  As consumers, users are legally required to take used batteries to appropriate collection  sites, the retail store where the batteries were purchased, or wherever batteries are sold.  Disposal: Do not dispose of this instrument in household waste. The user is obligated to take end‐of‐ life devices to a designated collection point for the disposal of electrical and electronic equipment.               EX36x‐en‐GB_V1.6   1/17 ...
  • Page 21 Fuse Replacement   WARNING: To avoid electrical shock, remove the test leads, disconnect the meter from  any circuit and turn OFF the meter before opening the case. Do not operate with an open case.  Follow the steps in the accompanying 11A/1000V (10x38mm) fuse replacement diagram.   Remove the test leads from the meter.  Remove the meter’s protective cover.  Remove the four (4) screws that secure the meter housing.  Replace the fuse with one of the same type and rating. Reassemble the meter before  use.  Fig 5‐2 FUSE REPLACEMENT                    Cleaning and Storage  Periodically wipe the case with a damp cloth and mild detergent; do not use abrasives or solvents.   LCD Diagnostic Test  Press and hold the H button while turning the meter ON to view all LCD icons and symbols. Turn the  meter OFF to end this diagnostic test. If there is reason to suspect that a display icon is not  functioning, this is a handy verification. If the display is responding unusually, or if an icon is  malfunctioning, please return the meter for service.  EX36x‐en‐GB_V1.6   1/17 ...
  • Page 22: Specifications

    6. Specifications  ELECTRICAL SPECIFICATIONS    Accuracy is given as ± (% of reading + least significant digit) at 23 C ±5 C with relative humidity  lower than 80%. Accuracy is specified for a period of one year after calibration.       1. Temperature Coefficient is 0.1 x specified accuracy /  C, < 18 C (64.5 F), > 28 C (82.4 F)  2. AC Functionality: ACV and ACA specifications are AC coupled, True RMS; the accuracy for  square waveforms is unspecified. For non‐sinusoidal waveforms, additional accuracy Crest  Factor (C.F.) considerations exist as detailed below:  Add 3.0% for C.F. 1.0~2.0; Add 5.0% for C.F. 2.0~2.5; Add 7.0% for C.F. 2.5~3.0  Function  Range  Resolution Accuracy (of reading) ‘OL’ Reading  mV DC 600.0mV  0.1mV  ± (0.5% + 5 digits)  660.0mV    Input Impedance: 10M; Overload Protection: AC/DC 1000V  6.000V  0.001V  6.600V ...
  • Page 23 Function  Range  Resolution Accuracy (of reading) ‘OL’ Reading  µA AC (EX363 only) 600.0 µA  0.1uA  ± (1.5% + 3 digits)  660.0µA    LCD Displays ‘0’ counts when the reading < 10 counts Input Impedance: 3kΩ approx.  Frequency Response: 45~500Hz (sine wave)  Over load protection: 1000V AC/DC  6.000 A  0.001    6.600A  AC Current (EX365) ± (1.5% + 3 digits)    10.00 A  0.01   20.00A  In 6A Range, the LCD displays 0 counts when the reading < 20 counts. In10A Range, the LCD displays 0 counts when the reading < 10 counts.  Maximum measurement time: >5A for max. 3 minutes with at least 20 minutes rest time.   >10A for max. 30 seconds with at least 10 minutes rest time.  Frequency Response : 45 ~ 500Hz (Sine Wave)  Overload Protection : AC/DC 11A   6.000 A  0.001    6.600A  DC Current (EX365) ± (1.0% + 3 digits)    10.00 A  0.01   20.00A  Maximum measurement time : >5A for max. 3 minutes with at least 20 minutes rest time. >10A for max. 30 seconds with at least 10 minutes rest time. ...
  • Page 24 Function  Range  Resolution Accuracy (of reading)  ‘OL’ Reading  100.00 Hz  0.01 Hz  100.00 Hz  1000.0 Hz  0.1 Hz  1000.0 Hz  Frequency ± (0.1% + 2 digits)    10.000 kHz  0.001kHz  10.000 kHz  100.00 kHz  0.01kHz  100.00 kHz  Minimum Sensitivity: > 5V (for ACV 1Hz ~ 10kHz) > 20.0V (for ACV 10kHz ~ 50kHz) unspecified (for ACV 50kHz ~ 100kHz) >0.6A (for ACA)  Minimum Frequency: 1Hz  Overload Protection: AC/DC 1000V or 11A ‐40.0~400.0  C    ± (1.0% + 20 digits)*  440.0  C  ‐40.0~752.0  F    ± (1.0% + 36 digits)*  824.0  F  *Does not include accuracy of the temperature probe. Accuracy specifications  Temp. (EX363)   assume surrounding temperature stable to ±1 C. For ambient temperature ...
  • Page 25 GENERAL SPECIFICATIONS  Display    6000‐count Multi‐Function LCD  Over‐range indication  “OL” or “‐OL” is displayed  Conversion rate  3 updates per second  Maximum Voltage  1000VAC RMS or 1000V DC maximum applied to any terminal  Polarity indication  Automatic: positive implied, negative indicated (‐)  Low battery indication      is displayed. Meter also displays available battery power at  startup  Auto Power OFF  After 20 minutes (can be disabled by holding the M (MODE) button  while turning the meter ON)  Operating Temperature and Humidity  ‐10~10C (14~50F)   10~30C (50~86F); 80%RH maximum   30~40C (86~104F); 75%RH maximum  40~50C (104~122F); 45%RH maximum  Storage Temperature and Humidity  ‐20~60C (‐4~140F); 80%RH maximum (with battery removed)   Operating Altitude  2000m (6562’)  Battery power    9V battery (battery life 200 hours typical with alkaline battery types)  Fuse protection  11A/1000V (10x38mm) AC/DC high energy (IR 20kA) DMM‐B‐11A         (EX365 only)  Weight   250g (8.8 oz.)Including battery  Dimensions (W x H x D)  74 x 156 x 44mm (2.9 x 6.1 x 1.7”) ...