Guidance and manufacture's declaration – electromagnetic

immunity for ME EQUIPMENT

and ME SYSTEMS that are not LIFE-SUPPORTING
I m m u n i t y
test
Conducted RF
IEC 61000-4-6
Radiated RF
IEC 61000-4-3
36
IEC
60601
compliance
test level
level
3 V
3 V
rms
150 kHz - 80 MHz
3 V/m
3 V/m
80 MHz - 2,5 GHz
Electromagnetic environ-
ment - guidance
Portable and mobile RF commu-
nications equipment should be
used no closer to any part of the
device, including cables, than
the recommended separation di-
stance calculated from the equa-
tion applicable to the frequency
of the transmitter.
rms
Recommended separation distance:
d= 1.167 √P
d= 1.167 √P 80 MHz - 800 MHz
d=2,333 √P 800 MH - 2,5 GHz
where P is the maximum output power
rating of the transmitter in watts (W)
according to the transmitter manufac-
turer and d is the recommended sepa-
ration distance in metres(m).
Field strengths from fixed RF
transmitters, as determined by
an electromagnetic site survey
,
pliance level in each frequency
range.
Interference may occur in the vi-
cinity of equipment marked with
the following symbol:
should be less than the com-
a
b