Guidance And Manufacture's Declaration - Electromagnetic Immunity For All Equipment & Systems - HMM smartLAB profi User Manual

Upper arm blood pressure monitor with wireless data transfer module
Table of Contents

Advertisement

Guidance and manufacture's declaration – electromagnetic
immunity for all equipment & systems
The smartLAB
environment specified below. The customer of the user of the device
should assure that it is used in such and environment.
Immunity test
Electrostatic
discharge (ESD)
IEC 61000-4-2
Electrical
fast
transient/burst
IEC 61000-4-4
Surge
IEC 61000-4-5
34
®
profi+ is intended for use in the electromagnetic
IEC 60601 test level
±6 kV contact
±8 kV air
±2 kV for power sup-
ply lines
± 1 kV line(s) to
line(s)
compliance level
±6 kV contact
±8 kV air
±2kV for power
supply lines
±1 kV differential
mode
Electromagnetic
environment
-
guidance
Floors should be
wood,
concrete
or ceramic tile. If
floor are covered
with
synthetic
material,
the
relative humidi-
ty should be at
least 30%.
Mains
power
quality should be
that of a typical
commercial
or
hospital environ-
ment.