Fuse Replacement - Agilent Technologies U1231A Service Manual

Handheld digital multimeter
Hide thumbs Also See for U1231A:
Table of Contents

Advertisement

Fuse Replacement

N O T E
C A U T I O N
U1231A/U1232A/U1233A Service Guide
No recalibration is required after replacing the fuse.
The current input terminals of your multimeter are fuse
protected. The fuses are located next to the battery
compartment.
The terminal is protected by a 10 × 38 mm, 11 A/1000 V,
30 kA fast- acting fuse (Fuse 1).
If you are certain that the fuse is faulty, replace it with one
of the same size and rating.
Before you proceed with the fuse replacement, remove all cable
connections to the terminals and ensure that the rotary switch is at the
OFF position.
1 Open the battery cover. Lift the tilt stand and loosen
screws with a suitable Phillips screwdriver and remove
the battery cover.
2 Locate the faulty fuse. The fuse is located at the bottom
of the batteries (see
Gently remove the defective fuse by prying one end of the
fuse with a flathead screwdriver and removing it out of
the fuse bracket. Replace a new fuse of the same size and
rating into the center of the fuse holder.
3 Close the batter cover. Place the battery cover back in its
original position and tighten the screws.
Service and Maintenance

Fuse Replacement

Figure 2-
2).