Battery Saver ("Sleep Mode"); Terminals; Rotary Switch Positions - Fluke 110 User Manual

True rms multimeters
Table of Contents

Advertisement

Models 110, 111 & 112
Users Manual

Battery Saver ("Sleep Mode")

The Meter automatically enters "Sleep mode" and blanks the
display if there is no function change or button press for 20
minutes. To disable the Sleep mode, hold down the Hz button
while turning the Meter on. The Sleep mode is always disabled in
the MIN MAX AVG mode.

Terminals

A
1
10 A
FUSED
2
Item
Input terminal for AC and DC current measurements to
1
10 A, or to 20 A overload for 30 seconds maximum
(Models 111 and 112 only), and for frequency of
current.
Common (return) terminal for all measurements.
2
Input terminal for voltage, continuity, resistance, diode
3
test, capacitance, and voltage frequency
measurements.
2
COM
V
3
aej01f.eps
Description

Rotary Switch Positions

Switch Position
AC voltage from 300 mV to 600 V.
K
Frequency from 5 Hz to 50 kHz.
Hz (button)
DC voltage from 1 mV to 600 V.
L
Frequency from 5 Hz to 50 kHz.
Hz (button)
Beeper turns on at
R
at
250
Ohms from 0.1
e
Diode test. Displays OL above 2.4 V.
G
Farads from 1 nF to 9999 F.
E
AC current from 3 A to 10 A.
(Models 111 & 112)
(20 A overload for 30 seconds maximum.)
?
10.00 display flashes.
20 A, OL is displayed.
Hz (button)
Frequency from 50 Hz to 5 kHz.
DC current from 0.001 A to 10 A.
(Models 111 & 112)
(20 A overload for 30 seconds maximum.)
A
10.00 display flashes.
20 A, OL is displayed.
Hz (button)
Frequency from 50 Hz to 5 kHz.
Notes: AC voltage and current AC-coupled, True RMS, up to 500 Hz.