Setting Up Logger for a P Type Job

If PQ Log is connected to the Logger, the Logger automatically identifies the Logger
type and the current probe type attached. To set up P type Logger Job:
1. Make sure the PC is communicating with the Logger. Starting from step 3 above,
set up the Logger parameters for the job as follows, starting from the top of the
"Set up for Logger" function P menu.
2. Enter the Nominal Voltage. This can be a phase-to-neutral voltage or a phase-to-
phase voltage.
3. Select 50 Hz or 60 Hz from the Frequency pull-down menu.
4. Select the Power Type from the pull-down menu. This is where you set up the
Logger for Delta (3-wire) or Wye (star or 4-wire), or other power configurations
from a list of choices.
5. Select the Min-Max-value from the pull-down list. This determines the averaging
period of min/max capture for rms voltage levels for later display. Some
standards require certain min/max averaging periods. The shortest setting is ½-
cycle, which is the most sensitive for measuring the absolute highs and lows that
occur within an overall averaging period for measurements. (You choose
averaging period for measurements in in Figure 2.)
6. Click on the Current button on the right side of the menu. The Current menu
appears.