Instrument Wiring For 4973 Cells With Quick Disconnect Cable; Wiring Model 4973 With Quick Disconnect Cable To Uda2182; Wiring Model 4973 With Quick Disconnect Cable To Junction Box; Figure 4-4 Wiring Diagram For 4973 Cells With Quick Disconnect Cable Connected To Uda2182 - Honeywell 04973 Series Installation And Maintenance Manual

Conductivity cells
Table of Contents

Advertisement

4.3 Instrument Wiring for 4973 Cells with Quick Disconnect Cable

4.3.1 Wiring Model 4973 with Quick Disconnect Cable to UDA2182

Figure 4-4 Wiring Diagram for 4973 Cells with Quick Disconnect Cable Connected to

4.3.2 Wiring Model 4973 with Quick Disconnect Cable to Junction Box

DRAWING 50012875
Figure 4-5 Installation diagram for 4973 cells with Quick Disconnect Cable connected to
10/07
Wire
Signal
Color
Name
Cell Low
Yellow
10
9
8
Coax
7
Cell High
6
RTH 3rd Wire
Jumper
5
Green
RTH Low
Red
4
RTH High
3
Brown
Blue
2
Wire to chassis
Earth Ground
1
ground screw
UDA2182 Analyzer
50
20 OR 50 FT. CABLE LENGTH.
junction box
Installation and Maintenance Manual
Electrical Connections
YELLOW
COAX
13