Specifications

2.1 Specifications for 04973 Series

Parameter
Cell Constant
Electrode Material
Maximum Pressure Limit
Maximum Continuous Temperature Limit
Materials of Construction
Cable Options
Approvals
Insertion
Insertion Depth

2.2 Specifications for 055919 Flow Chamber

Parameter
Maximum Flow
Maximum Pressure
Maximum Temperature
Dimensions
Materials of Construction
4
2. Specifications
04973 Series: 0.01, 0.1, 1.0 and 10 as specified
0.01 and 0.1 constant, titanium
1.0 and 10 constant, high density graphite with
Teflon sheath
1724 kPa (250 psig) at rated temperature
140°C (284°F) at rated pressure
For PVC wire:
105°C (221°F)
Cell Body: PES (polyethersulfone)
Electrodes: see 'Electrode Material'
Leadwire: PVC insulated 18 gage cable
20 and 50 feet lengths available
Quick Disconnect Option
Universal Head (Aluminum)
Manufactured to comply with ASME boiler and
pressure vessel code Section III, Div.1, UG-101
CRN #0F11607.5C
3/4" NPT male, Schedule 40 and 80 pipes
3-1/2" (89 mm) for 1, 10 and 0.01 cell constants
from solution end of 3/4" MNPT
2-1/2" (64mm) for 0.1 cell constant
2 gpm @ 40psig and atmospheric discharge
200 psig @ 25°C
140°C (284°F) at atmospheric pressure
See Figure 3-1
Polyethersulfone (PES)
Installation and Maintenance Manual
Description
Description