3.5 Torque Recommendations
For inserting a cell in metal fittings or bushings - 40 ft-lb maximum. For inserting a cell in plastic fittings
or bushings - 10 ft-lb maximum. Always use pipe sealant (preferably Teflon tape).
3.6 Dimension Drawings
1 1/ 8"
1 1/ 8"
1 1/2"
1 1/2"
(38mm)
(38mm)
(38mm)
(38mm)
Hexagon
Hexagon
1 1/2"
1 1/2"
(38mm)
(38mm)
Allow 4 1/8" (105mm) for re
Allow 4 1/8" (105mm) for re moval of cell
Notes:
1. Mount cell and flow chamber horizontally as shown above with flow exit up to eliminate
possible air gap around cell body.
2. If cell and flow chamber must be mounted vertically, attach a short length of tubing to flow
exit as shown below and form a trap to ensure filling of flow chamber, especially at low flow.
10/07
Flow Out
Flow Out
3/4" NPT (Female)
3/4" NPT (Female)
1 1/2"
1 1/2"
(38mm)
(38mm)
Octagon
Octagon
8 3/4"
8 3/4"
(222mm)
(222mm)
14 1/2" max
14 1/2" max
(368mm)
(368mm)
3/4" Fitting
3/4" Fitting
moval of cell
2" min.
(51mm)
Figure 3-1 Dimension Drawing for 055919 Flow Chamber
Installation and Maintenance Manual
Flow Chamber
Flow Chamber
Flow In
Flow In
3/4" NPT
3/4" NPT
(Male)
(Male)
Installation
a/n23342
7