Calibration Results; Calibration Pass; Calibration Fail - Honeywell MicroDock II Operating Manual

Automatic test and calibration station
Table of Contents

Advertisement

Calibration Results

After the calibration is complete, the station displays the results.
Pass
Fail
Gas type not detected (H
H2S –
all gases)
Sensor is disabled

Calibration Pass

and Pass indicates that the calibration has passed. The detector is now ready
for use. Press

Calibration Fail

If calibration fails,
If an inlet is not setup correctly during a calibration, the station displays the gas type
followed by a hyphen (–) and the result of the test as Fail.
This example displays H
If a test gas fails, verify that the
gas cylinder is not empty,
connections are attached correctly, and
inlets are setup correctly.
If an audible or visual test fails,
For information regarding solutions, refer to Troubleshooting.
54
OK to return to the normal operating screen.
displays beside the applicable gas (e.g., H2S
S, CO, and LEL as gases not detected.
2
displays beside the applicable test (AUD or VIS).
S is used as an example, this applies to
2
MicroDock II Automatic Test and Calibration Station
).