Calibration Results
After the calibration is complete, the station displays the results.
Pass
Fail
Gas type not detected (H
H2S –
all gases)
–
Sensor is disabled
Calibration Pass
and Pass indicates that the calibration has passed. The detector is now ready
for use. Press
Calibration Fail
If calibration fails,
If an inlet is not setup correctly during a calibration, the station displays the gas type
followed by a hyphen (–) and the result of the test as Fail.
This example displays H
If a test gas fails, verify that the
•
gas cylinder is not empty,
•
connections are attached correctly, and
•
inlets are setup correctly.
If an audible or visual test fails,
For information regarding solutions, refer to Troubleshooting.
54
OK to return to the normal operating screen.
displays beside the applicable gas (e.g., H2S
S, CO, and LEL as gases not detected.
2
displays beside the applicable test (AUD or VIS).
S is used as an example, this applies to
2
MicroDock II Automatic Test and Calibration Station
).