Honeywell MicroDock II Operating Manual page 48

Automatic test and calibration station
Table of Contents

Advertisement

If more than one docking module is queued: If more than one docking
module is queued for a bump check, the first module queued begins the
bump check.
The RUN LED flashes yellow on the remaining docking modules that are
queued for a bump check. When the first module is complete, the next
module in the queue begins the bump check.
4.
From
the docking module, press
yellow.
5.
Depending upon the type of detector, the option to reconfigure is
provided.
If the detector can be reconfigured, the following two screens display.
To reconfigure the detector, press
new configuration data to the detector. If required, refer to
the
Detector.
Or
Press
NOTe
If a selection is not made within 15 seconds, the station automatically
defaults to the No selection and proceeds with the bump check.
6.
The station begins the bump check and displays the
bay number,
process being performed,
type of detector, and
serial number of the detector.
NOTe
The X5 beeps and flashes twice to test the sensors when the bump check
begins.
7.
If there is a gas conflict or a selected gas cannot be found, the gas select
screen displays.
Gas conflicts must be resolved to continue the bump check. Refer to
Conflicts.
48
No to continue with the bump check.
BUMP CHECK. The RUN LED lights
Yes. The MMC card then transfers the
MicroDock II Automatic Test and Calibration Station
Reconfiguring
Gas

Hide quick links:

Advertisement

Table of Contents
loading