process being performed,
type of detector, and
serial number of the detector.
NOTe
The X5 beeps and flashes twice to test the sensors when calibration
begins.
X5 with CO
Sensor
2
NOTe
If calibrating a X5 detector, the station automatically zeroes and then
calibrates the CO
The station zeroes the CO
nitrogen (N
).
2
Ensure N
has been configured to an inlet. If required, refer to
2
station applies the gas.
The station then purges the N
The station then applies the gas to the CO
The station then begins calibrating the remaining sensors. After calibration is
complete, the station displays the results and performs a final purge to clear any
remaining gas.
The countdown of seconds remaining displays beside Purge.
NOTe
After calibration, the X5 automatically deactivates. To bypass the automatic
shutdown, remove the detector within 15 seconds after the final purge
begins.
MicroDock II Automatic Test and Calibration Station
sensor first.
2
sensor. The CO
2
before applying the CO
2
sensor can only be zeroed using
2
gas.
2
sensor.
2
Inlet
Setup. The
53