Using a small flathead screwdriver, press the release tab downwards. Pull
the adapter forward slightly and then lift upwards to remove.
4.
Hold the detector (LCD face up) at a 45° angle and insert the bottom into
the detector bay.
Ensure that the connector outlets on the bottom of the detector lock into
place over the connector pins in the docking bay, and then lower the top
into place.
5.
When the X5 displays "MICRODOCK", lower the lid and press down until
the release tabs click.
When the detector has been inserted correctly, the RUN LEDs on the docking
module light yellow and Unit Inserted displays on the station.
The station LCD displays the following docking module identification; the
type of detector that is inserted, and
bay and serial number of the docking module.
The X5 detector LCD displays Microdock.
6.
To bypass the station identification screen, press any
station.

Adding Another Docking Module

If required, refer to
module.
WArNINg
Only one module can be connected at a time. Complete steps #1-19 for each
docking module that is added.
To add another docking module, refer to Table 11 and 12, Figure 8, 9, and 10, and
complete the following procedures:
MicroDock II Automatic Test and Calibration Station
How to Use the Base Station
Housing Assembly
Wall mounting plate
Table 11. Torque Specifications
before adding a docking
Torque
9-10 in lbs
button on the
19