Status/Control Register; Channel Enable Register; Writing To The Registers - HP E1368A/69A/70A User Manual

Microwave switch and driver modules
Table of Contents

Advertisement

Status/Control
Register
Channel Enable
Register

Writing to the Registers

Channel Enable
Register
Note
Appendix B
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com
Each coaxial switch requires about 30 msec execution time (close to open or
open to close) during which time the switch is "busy". A read of the
Status/Control Register (base + 04
not busy or returns FF7F
16
A read of the Channel Register (base + 08
regardless of the channel states.
You can write the following Microwave Switch registers:
Channel Enable Register (base + 08
Writes to the Channel Enable Register (base + 08
close the desired channel. For example, write a "1" to bit 2 of the Channel
Enable Register to close channel 02. Or, write a "0" to bit of the register to
open channel 00.
All HP 3331X and HP 876X Microwave Switches are latching and will
remain in the last state during power down. Performing a reset (or at
power-up) opens all channels (refer to Figure 2-6). Any combination of
channel open or closed states is allowed.
) returns FFFF
16
16
(bit 7 is low-true) when the module is busy.
) always returns FFFF
16
)
16
) enable you to open or
16
HP E1368A/69A/70A Registers 77
when the module is
,
16