Power
Caution
Internal Power
External Power
Caution
24 Configuring the HP E1368A/69A/70A Modules
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com
After the coaxial switches have been installed (refer to " Installing Switches
on the HP E1369A" earlier in this chapter), the correct operating voltage
must be selected. See installed switch specifications.
The Microwave Switch modules are capable of providing +5V (set at
factory) or +12V using the mainframe power supply, or external voltages of
up to 42Vpk can be connected. Determine the required operating voltage
and proceed as instructed below. Refer to Figure 2-5 as required.
MAXIMUM CURRENT. The maximum current that the control
circuit can accommodate is 1 amp per switch. Maximum
current also depends on the output capacity of the mainframe
or power supply used.
Verify that the F3 fuse (P/N 2110-0712) is installed. Position jumpers as
instructed:
For 5V, jumper JM2 installed, and jumper JM151 removed.
For 12V, jumper JM151 installed, and jumper JM2 removed.
Verify that the F3 fuse (P/N 2110-0712) is removed (unsolder if necessary).
Connect the positive lead to EXT B+ on TB1, and the negative lead to EXT
GND on TB1. Connect leads to external power supply. Observe polarity.
MAXIMUM VOLTAGE. The maximum voltage that may be
applied to the EXTernal POWER terminal is 42Vpk.
Maximum wire size is No. 16 AWG. Wire ends should be stripped 6
mm (~ 0.25 in.) and tinned to prevent single strands from shorting
adjacent terminals.
It is recommended that each channel wire be identified (color coded
or marked) as the connection is not visible when the Microwave
Switch module is installed.
Verify that wires make good connections on screw terminals.
Chapter 2

Advertisement

Table of Contents
loading

This manual is also suitable for:

E1368aE1369aE1370a

Table of Contents