Frequency Sensitivity Specifications; Table 7-12 Frequency Sensitivity And Trigger Level Specifications For - Agilent Technologies U1251B User's And Service Manual

Handheld digital multimeter
Hide thumbs Also See for U1251B:
Table of Contents

Advertisement

7
Specifications
Table 7-11 Duty cycle and pulse width specifications
Function
Notes for duty cycle specifications:
1 The accuracy for duty cycle and pulse width is based on a 5 V square wave input into the DC 5 V range.
2 For AC coupling, the duty cycle range can be measured for signal frequency > 20 Hz.
Pulse width
Notes for pulse width specifications:
1 The accuracy for duty cycle and pulse width is based on a 5 V square wave input into the DC 5 V range.
2 Positive or negative pulse width must be greater than 10 µs and the range of duty cycle should be considered. The range
of pulse width is determined by the frequency of the signal.

Frequency Sensitivity Specifications

For voltage measurements
Table 7-12 Frequency sensitivity and trigger level specifications for voltage measurements
Minimum sensitivity (RMS sine wave)
Input
[1]
U1251B
range
20 Hz - 100
kHz
50.000 mV
10 mV
500.00 mV
25 mV
1000.0 mV
40 mV
5.0000 V
0.25 V
50.000 V
2.5 V
166
Mode
-
-
U1252B
>100 kHz -
20 Hz - 200
200 kHz
kHz
15 mV
10 mV
35 mV
70 mV
50 mV
120 mV
0.5 V
0.3 V
5 V
3 V
Range
Resolution
500 ms
2000 ms
Model Number
U1251B
>200 kHz -
< 100 kHz
500 kHz
25 mV
10 mV
150 mV
60 mV
300 mV
100 mV
0.5 V/1.25
1.2 V
V (< 100
Hz)
5 V
5 V
Agilent U1251B/U1252B User's and Service Guide
Accuracy (at full scale)
0.01 ms
0.2% + 3
0.1 ms
0.2% + 3
Trigger level for DC coupling
U1252B
>100 kHz -
< 100 kHz
200 kHz
15 mV
10 mV
70 mV
70 mV
150 mV
120 mV
0.6 V
0.6 V
6 V
6 V
> 100 kHz -
500 kHz
25 mV
150 mV
300 mV
1.5 V
15 V