Agilent U1251B/U1252B User's and Service Guide
To change the instrument calibration security code from the front
panel
1 When the multimeter is in the unsecured mode, press
for more than 1 second to enter the Calibration
Security Code setting mode.
2 The factory default calibration security code 1234 will be
displayed on the primary display.
3 Use the editing keys
character in the code.
4 Use the
and
the code.
5 Press
(Save) to store the new calibration security
code.
6 If the new calibration security code has been successfully
stored, the secondary display will show PASS.
Performance Tests and Calibration
and
to step each
to change each character in
6
137