Frequency Specifications - Agilent Technologies U1241B User's Manual And Service Manual

Handheld digital multimeters
Hide thumbs Also See for U1241B:
Table of Contents

Advertisement

6
Specifications and Characteristics

Frequency Specifications

Table 6-13 Freq
Function
[1]
Frequency
Frequency Sensitivity During Voltage Measurement
Frequency Sensitivity During Current Measurement
[1] The input signal is lower than the product of 20,000,000 V-Hz.
[2] For the accuracy of the maximum input, refer to AC voltage specifications.
[3] For the accuracy of the maximum input, refer to AC current specifications.
64
uency specifications with accuracy of ± (% of reading + No. of Least Significant
Digit)
Range
100.00 Hz
1000.0 Hz
10.000 kHz
100.00 kHz
1000.00 kHz
Input Range
(Maximum input for specified accuracy
= 10 x Range or 1000 V)
1000.0 mV
10.000 V
100.00 V
1000.0 V
Input Range
Minimum Sensitivity (R.M.S. Sine wave)
1000.0 μA
10000 μA
100.00 mA
440.0 mA
10.000 A
Resolution
Accuracy
0.01 Hz
0.03% + 3
0.1 Hz
0.001 kHz
0.01 kHz
0.1 kHz
Minimum Sensitivity (R.M.S. sine wave)
20 Hz to 50 kHz
0.3 V
0.5 V
5 V
50 V
20 Hz – 20 kHz
100 μA
500 μA
10 mA
50 mA
1 A
U1241B/U1242B User's and Service Guide
Min. Input Frequency
1 Hz
[2]
50 kHz to 200 kHz
0.6 V
1.8 V
10 V (<100 kHZ)
100 V (<100 kHZ)
[3]