Value
Preset
Reading room selection
This function is available only when the DICOM display function is
selected.
To select a reading room setting:
Select the DICOM display function. See To select a display
1.
function (DF):, on page 33.
In the ALC & DICOM Options menu, select Reading Room.
2.
Select the reading room that corresponds to the room type the
3.
display is installed in.
Return to the menu.
4.
You can select the following reading room types::
Name
DRR CR/DR/MAMMO
DRR CT/MR/NM
Description
The DICOM DF will be recalculated tak-
ing a preset ambient light value into
account. This value is determined by the
selected Reading Room and can be
adjusted manually within certain limits.
Description
Corresponds to light conditions in diag-
nostic reading rooms for computed radi-
ology, digital radiology or
mammography. This setting has the
lowest maximum ambient light.
Corresponds to light conditions in diag-
nostic reading rooms for computed
tomography, magnetic resonance or
nuclear medicine scans.
Display information